Azzota® Entry-level AAS Atomic Absorption Spectrometer, Flame/Graphite Furnace
-
Features
Specifications
| Spectrometer System | |
| Wavelength range | 190-900 nm |
| Light source | Hollow cathode lamp (HCL), Deuterium Arc lamp (D2) |
| Modulation | Square wave pulse |
| Modulation frequency |
100 Hz self reversal (SR) background, 400 Hz Deuterium Arc (D2) background |
| Monochromator | |
| Grating | 1800 grooves/mm diffraction grating |
| Blazed wavelength | 250 nm |
| Focus | 300 nm |
| Bandwidth | 0.1, 0.2, 0.5, 1.0, 2.0, 5.0 nm |
| Scan mode | Automatic |
| Photometric type | Single beam |
| Wavelength accuracy | ±0.15 nm |
| Wavelength resolution | 0.2 nm±0.02 nm |
| Wavelength reproducibility | <0.05 nm |
| Spectral bandwidth | 0.5, 1.0, 2.0, 5.0 nm |
| Baseline stability | 0.005 A/30 min |
| Background correction |
Deuterium Arc (D2) 1.0 Abs, Self Reversal (SR) 3.0 Abs |
| Flame Analysis | |
| Flame type | Air/acetylene |
| Sensitivity (Cu) | 2 μg/ml Absorption>0.280 Abs |
| Characteristic Concentration | Cu<0.02 μg/ml |
| Detection limit | Cu<0.004 μg/ml |
| Repeatability | Cu<0.7 % |
| Burner head | Titanium Alloy |
| Nebulizer |
High–efficiency glass, acid-proof available as an option |
| Pre-mix chamber | Corrosion-resistant |
| Atomizer selection | Manual (AA990F) Automatic changeover (AA990FG) |
| Safety features | Burner switch, flame sensor, gas leak sensor, low gas pressure sensor, drain trap sensor, power loss protection |
| Graphite Furnace Analysis | |
| Graphite head | Transversally heated |
| Temperature range | Ambient-2650 ℃ |
| Heating program | Up to 10 steps. Drying, Ashing, Atomization, Cleaning |
| Feedback | Voltage and optical temperature control feedback |
| Sensitivity (Cu) | 50 ng/m Absorption>0.40 Abs |
| Detection limit | Cd<0.004 ng/ml |
| Repeatability |
Cu<2.0%, Cd<2.0% |
| Graphite tubes | Pyrolytically coated with a platform |
| Sample size | Up to 20μl |
| Graphite cooling | Water circulator available |
| Safety features |
Burner Switch, Flame Sensor, Gas leak Sensor, Drain Trap Sensor, Power Loss Protection, Circulation Water (graphite), Over Temperature Sensor (graphite) |
| Data Processing | |
| Operating program | AA Win software |
| Analytical methods |
Flame AA, Flame AE, Graphite Furnace, Hydride Generation |
| Readout Mode | Continuous, manual, peak height, peak area |
| Calibrations | Multi-standard calibration, standard addition, interpolation |
| Data storage | Analytical results, instrument and measurement parameters, signal profile, calibration curve |
| Power Requirements | |
| Main unit | 220 VAC 50/60 Hz 200 W |
| Graphite power supply |
380 VAC 50/60 Hz instantaneous power 5kW current rating 7 kW |
| Dimensions | |
| Main unit |
44× 20 × 20 inch 158 lb
|
| Graphite power supply |
20 × 20 × 18 inch 154 lb |